Penetration testing for medical devices and IVDs

Penetration testing for medical devices, filed as evidence

A reference for regulatory affairs and product security leads. Which requirements and guidance may be relevant to a connected device, what testing can demonstrate, and where the evidence belongs in technical documentation or a premarket submission.

  • MDR Annex I 17.2
  • IVDR Annex I 16.2
  • MDCG 2019-16 §3.7
  • FD&C 524B
  • ANSI/ISA 62443-4-1
  • ENISA GENERAL-11.01

Part 01 Scope and definitions

What a medical device penetration test is

A time-boxed, manual attack on one device build, the software that ships with it and the services it talks to, carried out by people who did not design it.

A medical device penetration test starts from a defined build: a firmware version, a companion application version, a backend release and the accounts and interfaces a real user or a real attacker can reach. Testers work through those interfaces the way an adversary would, chain what they find, and record what the device does about it. The output is a report with a scope, a duration, a method and a set of findings, each with the evidence needed to reproduce it.

The regulatory frame is short. MDR Annex I, point 17.2 requires software in a device to be “developed and manufactured in accordance with the state of the art taking into account the principles of development life cycle, risk management, including information security, verification and validation”. IVDR Annex I point 16.2 is the same sentence. Two lines of law import whatever the state of the art happens to be, and MDCG 2019-16 rev.1 is where the Medical Device Coordination Group writes down what that means in practice.

Chapter 3.7 of that guidance is the sentence to know: “The primary means of security verification and validation is testing. Methods can include security feature testing, fuzz testing, vulnerability scanning and penetration testing.” A footnote defines penetration testing, via ENISA, as “the assessment of the security of a system against different types of attacks performed by an authorised security expert”. Chapter 4.1 then files it: Annex II documentation “shall comprise a justification, validation and verification of the solutions adopted to meet those requirements (e.g. methods and results of security testing described in Chapter 3.7 of this guidance)”.

On the US side the requirement is statutory rather than interpretive. Section 524B of the FD&C Act, added by the Food and Drug Omnibus Reform Act of 2022 and effective 29 March 2023, makes cybersecurity information part of the content of any 510(k), PMA, PDP, De Novo or HDE for a cyber device. The FDA guidance issued on 3 February 2026 says penetration test reports “should be provided”, lists the five elements they must contain, and puts tester independence first on that list.

Part 02 Requirements and guidance

The requirements behind test evidence

Eleven entries drawn from regulations and guidance. Their legal effect and applicability differ; each identifies its source.

  1. MDR I · 17.2

    State-of-the-art software development, including information security

    “For devices that incorporate software or for software that are devices in themselves, the software shall be developed and manufactured in accordance with the state of the art taking into account the principles of development life cycle, risk management, including information security, verification and validation.” Regulation (EU) 2017/745, consolidated 10 January 2025. IVDR Annex I point 16.2 is the mirror text.

    Applies per device · assessed at conformity assessment and on significant change

  2. MDR I · 17.4

    Minimum hardware, IT network and IT security requirements

    “Manufacturers shall set out minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.” The same words appear at IVDR Annex I point 16.4. MDCG 2019-16 §3.6 adds that these requirements are the manufacturer’s to determine, and that they must not become a way of compensating for the device’s own weaknesses.

    Applies per device · published in the instructions for use

  3. MDCG §3.7

    Testing as the primary means of security verification and validation

    “The primary means of security verification and validation is testing. Methods can include security feature testing, fuzz testing, vulnerability scanning and penetration testing.” MDCG 2019-16 rev.1, chapter 3.7. The accompanying footnote takes the ENISA definition of penetration testing.

    Guidance · what a notified body works from

  4. MDCG §4.1

    Security testing results belong in the Annex II documentation

    Documentation “shall comprise a justification, validation and verification of the solutions adopted to meet those requirements (e.g. methods and results of security testing described in Chapter 3.7 of this guidance)”, and “needs to be updated with information raised through the manufacturers post market surveillance system related to handling and remediation of cybersecurity incidents and vulnerabilities”.

    Guidance · the filing instruction

  5. MDR II · 6

    Results and critical analyses of all verification and validation tests

    Annex II section 6: “The documentation shall contain the results and critical analyses of all verifications and validation tests and/or studies undertaken to demonstrate conformity of the device with the requirements of this Regulation.” Section 6.1(b) names software verification and validation specifically, including all hardware configurations and operating systems the manufacturer identifies.

    Applies per device · the technical documentation itself

  6. MDR Art. 83(3)

    The file is kept current from post-market surveillance

    Article 83(1) makes post-market surveillance part of the quality management system. Article 83(3) lists what the data are used for and closes with the sentence that governs a security file: “The technical documentation shall be updated accordingly.”

    Continuing · for the life of the device

  7. 524B(a)

    Cybersecurity information is part of the submission

    A person who submits a 510(k), PMA, PDP, De Novo or HDE for a device meeting the section 524B(c) definition of a cyber device must submit information to ensure the device meets the requirements in 524B(b). The definition has three limbs: software validated, installed or authorised by the sponsor; the ability to connect to the internet; and technological characteristics that could be vulnerable to cybersecurity threats.

    Statutory · effective 29 March 2023 · per submission, including modifications

  8. 524B(b)(3)

    An SBOM, including commercial, open-source and off-the-shelf components

    The FDA recommends machine-readable SBOMs consistent with the minimum elements identified in the October 2021 NTIA “Framing Software Component Transparency” document, plus, for each component, the level of support provided through monitoring and maintenance and the component’s end-of-support date.

    Statutory · per submission

  9. FDA V.C

    A penetration test report with five named elements

    “Penetration test reports should be provided and include the following elements: Independence and technical expertise of testers; Scope of testing; Duration of testing; Testing methods employed; and Test results, findings, and observations.” Cybersecurity in Medical Devices, FDA, 3 February 2026.

    Guidance · per submission

  10. FDA V.C

    Post-release testing at regular intervals, for example annually

    “After release, cybersecurity testing should be performed at regular intervals commensurate with the risk (e.g., annually) to ensure that potential vulnerabilities are identified and able to be addressed prior to their ability to be exploited.”

    Guidance · recurring for the marketed lifetime

  11. ENISA 11.01

    Hospitals now ask for the evidence at procurement

    Control GENERAL-11.01 in ENISA’s procurement guidelines for hospitals and healthcare providers (July 2026), marked Level 1 and applying to all procurement service types: “Confirm that security validation (e.g., penetration testing, audits, independent assessments) is performed on procured products or services before deployment.” The party who produces that evidence is the vendor.

    Buyer-side guidance · per tender

The MDR and IVDR entries are quoted from the consolidated texts of 10 January 2025 (02017R0745 and 02017R0746, revision 005.001). Guidance entries are quoted from MDCG 2019-16 rev.1 (December 2019, July 2020 rev.1) and the FDA guidance issued 3 February 2026. This is a description of published requirements, not legal advice on your device.

Part 03 Interactive · rules from the cited instruments

Build the cybersecurity evidence file for your device

Set the markets and device characteristics. The panel suggests requirements to assess, test evidence to prepare, and where to file it. Confirm legal scope and device classification before treating an entry as applicable.

Interactive mode is not available. You can read the full reference content below. No answers are assessed and no result is calculated.

  • Placing the device on the EU market. MDR Annex I points 17.2 and 17.4 apply, with IVDR points 16.2 and 16.4 as the mirror text. MDCG 2019-16 rev.1 §3.7 names penetration testing among the methods of security verification and validation, and §4.1 files the methods and results in the Annex II technical documentation. MDR Annex II section 6 requires the results and critical analyses of all verification and validation tests, and Article 83(3) requires the documentation to be updated as post-market surveillance produces new information.
  • Submitting to the FDA. Section 524B of the FD&C Act applies to any 510(k), PMA, PDP, De Novo or HDE covering a cyber device. The submission carries a vulnerability plan under 524B(b)(1), processes under 524B(b)(2) and an SBOM under 524B(b)(3). The February 2026 guidance asks for a penetration test report containing the independence and technical expertise of testers, the scope, the duration, the methods employed, and the results, findings and observations, and for the original third-party report where one exists.
  • Wireless, wired or cloud connectivity. Radio and network interfaces bring pairing, key exchange, replay, unauthenticated commands and listening services into scope, and make the FDA’s multi-patient harm view substantive. A manufacturer-hosted backend brings related systems in as well: the FDA counts update servers and connections to healthcare facility networks among them.
  • A field update mechanism. The updateability and patchability view has to describe the end-to-end deployment path, including technology the manufacturer does not control. Testing covers signature and version verification, rollback and downgrade behaviour, and the server-to-device path.
  • Patient data on the device. MDCG 2019-16 Table 3 lists personal data storage confidentiality, personal data integrity and authenticity, and de-identification among the indicative security capabilities. Testing covers protection at rest, key storage, factory reset and what can be recovered from a decommissioned unit.
  • Selling into an EU hospital. ENISA’s July 2026 procurement guidelines make independent security validation before deployment a Level 1 control for all procurement service types, and ask buyers to request an MDS2 form from device suppliers.
  • After release. The FDA expects testing at regular intervals commensurate with the risk, for example annually. In the EU the trigger is significant change plus whatever post-market surveillance produces, because the technical documentation has to be updated accordingly.
Scope a device test

Markets you place the device on

Device characteristics

Cybersecurity evidence file19 entries

Requirements to assess

  1. MDR I · 17.2 State-of-the-art software development, including information security Development life cycle, risk management, verification and validation. IVDR Annex I point 16.2 is identical.
  2. MDR I · 17.4 Minimum hardware, IT network and IT security requirements Set by you, published in the instructions for use, and not a way of compensating for weaknesses in the device.
  3. MDCG §3.7 Security verification and validation by testing Penetration testing is one of the named methods, alongside security feature testing, fuzz testing and vulnerability scanning.
  4. MDCG §4.1 Methods and results of security testing in the documentation Filed with the justification, validation and verification of the solutions adopted, and kept updated from post-market surveillance.
  5. MDR II · 6 Results and critical analyses of all verification and validation tests Including software verification and validation across every hardware configuration and operating system you identify.
  6. MDR Art. 83(3) Technical documentation updated from post-market surveillance A finding after release is a reason to revise the file, not only the product.
  7. NIS2 21(2)(e) Security in acquisition, development and maintenance Device and IVD manufacturing is listed in NIS2 Annex II point 5(a). Applicability also depends on Article 2 scope rules and national implementation; placing a device on the EU market alone does not establish it. In-scope entities must assess Article 21 measures, including vulnerability handling and supply-chain security.
  8. ENISA 11.01 Independent security validation before deployment An ENISA Level 1 procurement recommendation covering all procurement service types. It may become a buyer requirement when adopted in a tender; the guidance alone does not impose a legal duty on every hospital or supplier.

Test evidence to prepare

  1. TEST Independent penetration test of the defined build One report covering the device, its companion software and the services it talks to, with the tester’s independence from the development team stated on the front page.
  2. TEST · RF Radio interface testing Pairing and bonding, key exchange, replay and injection on the air interface, unauthenticated commands, and what the device accepts while unpaired or in service mode.
  3. TEST · UPD End-to-end update path testing Signature and version verification on the device, rollback and downgrade behaviour, and the whole path from the update server to the device, including technology you do not control.
  4. TEST · CODE Secure code analysis and open-source component scanning MDCG 2019-16 §3.7: “Additional security testing can be done by using tools for secure code analysis and tools that scan for open source code and libraries used in the product, to identify components with known issues.”
  5. ANALYSIS Security risk management output the test can be traced to MDCG 2019-16 §3.2 keeps security risk analysis, evaluation, control and residual-risk evaluation inside the device risk management process. Findings have to land back in it.
  6. ANALYSIS Verification of the minimum IT requirements you publish The operating-environment assumptions you state under 17.4 are testable claims. MDCG §3.6 says reliance on the environment should be kept to a minimum and documented where it exists.
  7. CADENCE Retest on significant change, and whenever surveillance says so Article 83(3) makes the technical documentation follow the post-market surveillance data, so a new attack path against your component stack is a file event.

Where the evidence is filed

  1. FILE · EU Annex II technical documentation, section 6 Results and critical analyses of all verification and validation tests, with the security testing methods and results from MDCG §3.7.
  2. FILE · EU Instructions for use The minimum hardware, IT network characteristics and IT security measures needed to run the software as intended, under MDR 23.4(ab) and IVDR 20.4.1(ah).
  3. FILE · EU Post-market surveillance file Handling and remediation of cybersecurity incidents and vulnerabilities, feeding revisions back into the technical documentation.
  4. FILE · EU Tender pack for hospital buyers Current independent security validation for ENISA GENERAL-11.01, and an MDS2 form, which PLAN-02.01 names as the best-practice request at procurement.
Scope a device test

This panel reproduces published requirements and links each source. It is not a regulatory determination: device classification, the conformity assessment route and the submission pathway are for you and your notified body or reviewer to establish.

Part 04 EU conformity file · US premarket submission

One test campaign, two evidence packs

The technical work behind a CE-marking file and an FDA submission is largely the same. What differs is what has to be written down, and how strictly independence is treated.

The same evidence, filed two ways
EvidenceEU · notified bodyUS · FDA 524B
Penetration test reportMDCG 2019-16 §3.7 names penetration testing as a method of security verification and validation; §4.1 files the methods and results in Annex II“Penetration test reports should be provided” with five named elements (guidance section V.C)
Tester independenceNot stated as a requirement in the MDR or in MDCG 2019-16The first required report element; third parties may be necessary, and the original third-party report must be provided
Scope, duration, methods, resultsAnnex II section 6: results and critical analyses of all verification and validation testsNamed individually as report elements
Threat model and architectureSecurity risk management under MDCG §3.2 and §3.4; no prescribed set of viewsGlobal system, multi-patient harm, updateability and patchability, and security use case views
SBOMNot an Annex I requirement; component control sits inside state of the art and supplier managementStatutory under section 524B(b)(3), machine-readable, with support level and end-of-support date
Detailed vulnerability testingMDCG §3.7 mentions fuzz testing, vulnerability scanning, secure code analysis and open-source component scanningEnumerated against ANSI/ISA 62443-4-1, from abuse cases to binary composition analysis
Operating environmentMDR 17.4 minimum IT requirements, plus the MDCG §3.6 rule against compensating for device weaknesses in the environmentCarried by the security use case views and the labeling
Retest cadenceSignificant change, plus whatever post-market surveillance produces: “The technical documentation shall be updated accordingly”“At regular intervals commensurate with the risk (e.g., annually)”
Sources: Regulation (EU) 2017/745, consolidated 10 January 2025; MDCG 2019-16 rev.1; Cybersecurity in Medical Devices, FDA, 3 February 2026.

Only one side of this table has a hard rule on who runs the test. The FDA guidance says manufacturers “should indicate in the test reports by whom the testing was performed (e.g., independent internal testers, external testers) and what level of independence those responsible for testing devices have from the developers responsible for designing devices”, and adds that “in some cases, it may be necessary to use third parties to ensure an appropriate level of independence between the two groups”. Where a third party is used, “manufacturers should provide the original third-party report”.

The MDR says nothing about who performs the test. In practice a notified body reads an internal report against MDCG 2019-16 §3.7 and asks the same questions the FDA writes down: what was in scope, how long it ran, what methods were used, and who could have been marking their own homework. Commissioning one independent campaign and reporting it twice costs less than doing the work once for Europe and again for a submission.

Sequence matters more than volume. Run the campaign against a release candidate, close the findings you intend to close, retest the fixes, then freeze the report. A report that lists open critical findings with no assessment is worse than no report at all: the FDA asks for “your assessment of any findings including rationales for not implementing or deferring any findings to future releases”, and the same expectation reaches a notified body through the risk file.

Part 05 Scope of a connected-device campaign

What actually gets tested

A device is not one target. These are the surfaces a campaign is scoped across, and the capability from MDCG 2019-16 Table 3 that each one is there to exercise.

  1. SURFACE 1

    The radio interface

    Bluetooth Low Energy, Wi-Fi, cellular or a proprietary protocol. Pairing and bonding, key derivation and storage, replay and injection, what the device accepts while unpaired, and whether a service or engineering mode is reachable over the air. Node authentication, transmission confidentiality, transmission integrity.

  2. SURFACE 2

    Wired network services

    Every listening port on the device, the authentication on each, transport protection, and the DICOM, HL7 or vendor protocols in between. ENISA control GENERAL-06.02 asks hospital buyers to confirm that device networks are isolated or on dedicated segments, so testing from a shared segment reflects what they will ask about. Node authentication, system and OS hardening.

  3. SURFACE 3

    The update path, end to end

    Signature and version verification on the device, rollback and downgrade behaviour, and the whole route from the update server to the unit. The FDA is explicit that this path “will likely include traversing technology that the device manufacturer does not control”. Cybersecurity product upgrades, configuration of security features.

  4. SURFACE 4

    Data at rest and the decommissioned unit

    Storage of patient data and of keys, what a factory reset actually removes, what a service technician can extract, and what remains on removable media or a returned device. Personal data storage confidentiality, personal data integrity and authenticity, personal data de-identification.

  5. SURFACE 5

    The companion application

    Mobile or desktop software that ships with the device: local storage, credential handling, certificate validation, the pairing flow, and the protocol it speaks to the device. MDCG 2019-16 §3.6 treats anything you did not supply as the operating environment, which makes the boundary a documentation decision as much as a technical one. Person authentication, automatic logoff, emergency access.

  6. SURFACE 6

    The manufacturer backend

    Clinician portal, device-management service, telemetry ingestion and the update server. Device-to-cloud authentication, tenant and account isolation, object-level authorization, and administrative interfaces. The FDA treats update servers and healthcare-facility network connections as related systems under 524B(b)(2). Authorization, audit controls, data backup and disaster recovery.

  7. SURFACE 7

    Service and maintenance access

    Debug headers, service menus, maintenance accounts, remote support tooling and the credentials that go with them. ENISA control GENERAL-06.08 puts strictly controlled remote maintenance access to medical devices on the buyer’s checklist. Physical locks, configuration of security features, audit controls.

The capability names in italics are quoted from Table 3, “Indicative list of security Capabilities for MD”, in MDCG 2019-16 rev.1 §3.3. The table is indicative, not a checklist: which capabilities apply is an output of your security risk management.

Part 06 FDA guidance section V.C · MDR Annex II section 6

What has to be in the report

Five elements are named by the FDA. Two more are what a notified body needs in order to accept the report as verification and validation evidence.

  1. ELEMENT 1

    Independence and technical expertise of testers

    Named first. The guidance asks manufacturers to state “by whom the testing was performed (e.g., independent internal testers, external testers) and what level of independence those responsible for testing devices have from the developers responsible for designing devices”, and notes that “in some cases, it may be necessary to use third parties to ensure an appropriate level of independence between the two groups”.

  2. ELEMENT 2

    Scope of testing

    Which build, which interfaces, which accounts and roles, which backend environment, and what was excluded and why. The scope statement is what a reviewer compares against your own architecture views.

  3. ELEMENT 3

    Duration of testing

    Calendar window and tester effort. A three-day review of a networked infusion platform reads differently from a three-week campaign against the same device, and the report should let a reviewer see which one happened.

  4. ELEMENT 4

    Testing methods employed

    The techniques actually used, including tooling. Footnote 47 of the guidance asks for “the name of the tool, version information as applicable, and any settings or configuration options for the tools used”.

  5. ELEMENT 5

    Test results, findings and observations

    Every finding with the evidence to reproduce it, plus, separately, “your assessment of any findings including rationales for not implementing or deferring any findings to future releases”. Deferred issues need a release plan: which vulnerabilities, when, whether interim devices receive the update and how long it takes to reach them.

  6. EU · A

    A traceable link back to the security risk file

    MDCG 2019-16 §3.2 keeps security risk analysis, evaluation and control inside the device risk management process. A report that cannot be traced to the risk controls it exercised is hard to file as verification of those controls under Annex II section 6.

  7. EU · B

    The original report, unedited

    The FDA requires the original third-party report where a third party performed the work. Keeping the same document in the EU technical file, rather than a summary written afterwards, means one artifact answers both reviewers.

The five named elements are quoted verbatim from Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, FDA, 3 February 2026, section V.C. The document is marked “Contains Nonbinding Recommendations” on every page; section 524B of the FD&C Act is the binding requirement it supports.

Part 07 ENISA · health sector

Why the buyer side is asking

Hospital procurement did not tighten in the abstract. These are the figures ENISA publishes for the sector your devices are sold into.

of analysed health-sector incidents were ransomware
54%
Jan 2021 – Mar 2023
of analysed incidents affected healthcare providers, in particular hospitals
53%
Jan 2021 – Mar 2023
of health-related incidents in the 2024 threat landscape were ransomware
45%
ENISA Threat Landscape 2024
of those health-related incidents were data breaches
28%
ENISA Threat Landscape 2024

Figures as published by ENISA on its health sector page, with the periods it states. ENISA also records that health is the most affected sector for four years in a row (2020-2023) in significant NIS incidents reported through CIRAS. Source: ENISA, Health

Part 09 Frequently asked

Questions manufacturers ask first

Short answers with the citation attached. Where a source does not settle a question, the answer says so.

Does the MDR require a penetration test?

Not by name. MDR Annex I point 17.2 requires software to be developed in accordance with the state of the art including information security, verification and validation, and Annex II section 6 requires the results and critical analyses of all verification and validation tests in the technical documentation. MDCG 2019-16 rev.1 §3.7 is where penetration testing is named, as one method among security feature testing, fuzz testing and vulnerability scanning, and §4.1 places the methods and results of that testing in the Annex II documentation. Guidance is not law, but it is what a notified body works from.

Does the FDA require an external tester?

It requires the report to state the independence and technical expertise of the testers, and says that “in some cases, it may be necessary to use third parties to ensure an appropriate level of independence between the two groups”. Where a third party is used, the original third-party report must be provided. So the rule is not “external always”; it is that you declare the level of independence, and that a low one has to be defensible for a device of that risk.

How often do we have to retest?

For the US market the guidance is explicit: “After release, cybersecurity testing should be performed at regular intervals commensurate with the risk (e.g., annually).” For the EU there is no stated interval. The trigger is significant change plus post-market surveillance, because MDR Article 83(3) ends with “The technical documentation shall be updated accordingly.” In practice most manufacturers selling into both markets run one annual campaign and use it for both.

Is our device a “cyber device” under section 524B?

Section 524B(c) defines it as a device that “(1) includes software validated, installed, or authorized by the sponsor as a device or in a device; (2) has the ability to connect to the internet; and (3) contains any such technological characteristics validated, installed, or authorized by the sponsor that could be vulnerable to cybersecurity threats”. All three limbs, in the statutory wording. Internet capability is the one that most often decides it, and it is capability rather than intended use.

Does the Cyber Resilience Act apply to our device?

No. Article 2(2) of Regulation (EU) 2024/2847 states that the regulation “does not apply to products with digital elements to which the following Union legal acts apply: (a) Regulation (EU) 2017/745; (b) Regulation (EU) 2017/746; (c) Regulation (EU) 2019/2144.” Devices and IVDs are out. Two things still reach you: products in your catalogue that are not devices, and the CRA-grade expectations your own suppliers will start meeting. Article 2(5) also lets the Commission adjust the boundary by delegated act, so it is worth rechecking rather than assuming it is settled forever.

Is EN IEC 81001-5-1 mandatory?

It is not cited in the Official Journal as a harmonised standard under Regulation (EU) 2017/745, so it carries no presumption of conformity. We checked that against Commission Implementing Decision (EU) 2021/1182 and all eleven amendments published to date, the most recent on 11 June 2026: none of them references 81001. That does not make secure-lifecycle evidence optional, because Annex I point 17.2 imports the state of the art regardless of what has been cited.

Are device manufacturers essential or important entities under NIS2?

Manufacturers of medical devices and IVDs are listed in Annex II point 5(a) of Directive (EU) 2022/2555, which is the “other critical sectors” annex, with an exception for makers of devices considered critical during a public health emergency, who sit in Annex I. Most commentary gets this backwards. Article 21(2)(d) and (e) are the measures that touch a product security programme directly: supply chain security, and security in acquisition, development and maintenance including vulnerability handling and disclosure.

What will a hospital ask us for before it buys?

ENISA’s procurement guidelines for hospitals and healthcare providers, published in July 2026, make GENERAL-11.01 a Level 1 control for all procurement service types: confirm that security validation, expressly including penetration testing, audits and independent assessments, is performed on procured products before deployment. Alongside it, PLAN-02.01 names the MDS2 form as best practice to request from device suppliers, PLAN-02.02 asks for evidence of following accepted guidelines and standards, and SOURCE-06.01 asks for contract clauses on timely vulnerability disclosure and patching.

What does a device penetration test cost?

This site does not publish a price, because a number without a scope is not information. The cost drivers are countable: the number of physical and radio interfaces, whether firmware has to be extracted and analysed, how many roles and tenants exist in the backend, whether a companion application is in scope, how many device units and spare parts you can supply, and how much retest is needed after remediation. A scoping call that walks those seven items produces a defensible figure; a rate card does not.

Can we reuse last year’s report for a new submission?

Only if the build has not moved. The guidance is clear that a modification submitted through one of the enumerated pathways also has to meet section 524B, and that changes which may impact cybersecurity “could include changes to authentication or encryption algorithms, new connectivity features, or changing software update process/mechanisms”. A report against a superseded firmware version is evidence about a device you no longer sell.